Accutron CT-D - mobile precision - double head technology - wireless
Technical data
Application
- Wireless, battery operated, fully digital, user programmable double-head injector for injection of contrast agent and physiological saline solution in the field of computer tomography
Injector concept
- Microprocessor controlled double-head injector with electric drive and two injection units
- Multiphase, program controlled injection of contrast agent and physiological saline solution
- Aluminium housing
- Integrated heatable syringe holders for Easy Loading Syringe (ELS) 200 ml
- Keep-Vein-Open (KVO) feature
- Motor, control and rechargeable batteries are integrated into a single compact unit
- Touch-screen control panel with different languages
- Alternatively, display of injection parameters or pressure graph
- Hand switch to start and stop the injection
- Buttons for manual piston movement, starting and stopping the injection
- Automatic power saving mode (stand-by-function)
Mechanical design
Detachable, swivelling injection unit on a mobile column stand
Secured injection position: injection position at an angle of about 15° below horizontal (built-in sensor), tilt vertically upwards to vent the system Total weight: 26 kg Power supply
- Operation is independent of the mains supply, thanks to high output rechargeable batteries with a capacity sufficient for up to two days' operation
- Input voltage charger: 100 – 240 V, 50 – 60 Hz
- Power consumption charger: < 100 VA
Memory capacity
- 80 injection profiles can be defined and stored by the user
Filling of injection syringes
- Automatic filling via menu with volume input or
Manual filling with variable speed- Filling speed: 1 to 5 ml/s
- Optimised high pressure tube systems with automatic valves
Injection parameters
Maximum injection volume
2 x 200 mlPartial injection volume
- 1 to 200 ml, programmable in steps of 1 ml
Maximum pressure
- 21 bar (304 psi), programmable from 5 to 21 bar in increments of 1 bar (72.5 to 304 psi in 14.5 increments)
Flow rate of the two injection units
- 0.1 – 10 ml/s, programmable in steps of 0.1 s
- Alternatively, input of flow rate or phase duration
Number of phases
- 1 to 6
Injection delay
0 – 255 s
Phase delay
- 0 – 255 s
Scan delay
- 0 – 255 s
Remote control
- Wireless touch-screen remote control
- Alternatively, display of injection parameters or pressure graph
Options
- Use of prefilled syringes
- CANopen Interface
- Ceiling suspension