Accutron HP-D - the only double head injector for angiography

Technical data

Application

  • Wireless, battery operated, fully digital user programmable double-head injector for injection of contrast agent and physiological saline solution in the field of angiography, intervention and computed tomography

Injector concept

  • Microprocessor controlled double-head injector with electric drive and two injection units
  • Multiphase, program controlled injection of contrast agent and physiological saline solution
  • Alternatively, single- or multi-injection mode
  • Aluminium housing
  • Integrated, heated syringe holders for Easy Loading Syringe (ELS) 200 ml
  • Motor, control and rechargeable batteries are integrated into a single compact unit
  • 10,4“ Touch-screen control panel with different languages
  • Buttons for manual piston movement (filling and venting)
  • Alternatively, display of injection parameters or pressure graph
  • Automatic power saving mode

Mechanical design

  • Detachable, swivelling injection unit on a mobile column
  • Secured injection position at an angle of about 15° below horizontal (built-in sensor), tilt vertically upwards to vent the system
  • Hand switch to control the injection
  • Total weight: 48 kg

Power supply

  • Operation is independent of the mains supply, thanks to high output rechargeable batteries with a capacity sufficient for up to two days' operation
  • Input voltage charger: 100 – 240 V, 50 – 60 Hz,
  • Power consumption charger: < 100 VA

Memory capacity

  • 60 injection profiles can be defined and stored by the user

Filling of injection syringes

  • Automatic filling via menu with volume input or
    Manual filling with variable speed
  • Filling speed 1 to 8 ml/s
  • Optimised high pressure tube systems with check valves

Injection parameters

Maximum injection volume

  • 2 x 200 ml

Partial injection volume

  • 1 – 200 ml, programmable in 1 ml increments

Maximum injection pressure

  • 83 bar (1200 psi), programmable from 5 to 83 bar in 1 bar increments (72.5 to 1200 psi in 14.5 psi increments)

Flow rate of the two injection units

  • 0.1 – 30 ml/s, programmable in 0.1 ml/s increments
  • Alternatively, input of flow rate or phase duration

Rise time

  • 0.1 – 10 s, programmable in 0.1 s increments

Number of phases

  • 1 – 3

Injection delay

  • 0 – 255 s


Phase delay

  • 0 – 255 s

X-ray delay

  • 0 – 255 s

Options

  • Wireless touch-screen remote control
  • Interface

Product information and technical data

Product Image

Subject to technical alterations