Accutron MR - Mobile precision for magnet resonance imaging

Technical data
Application
Wireless, battery operated, fully digital user programmable double-head injector for contrast agent and physiological saline solution in the field of magnetic resonance imagingInjector concept
Microprocessor controlled double-head injector with electric drive and two independent injection units Multiphase, program controlled injection of contrast agent and physiological saline solution Aluminium housing Motor, control and rechargeable batteries are integrated into a single compact unit Touch-screen control panel with different languages Additional keypad for manual piston movement, starting and stopping the injection Syringe holder with Easy Loading Syringe (ELS) 65 ml/200 ml "Keep-Vein-Open" (KVO) feature Automatic power saving mode Non-stick syringe holderMechanical design
Swivelling injection units connected to a mobile support with a column stand Operational angle of about 15° below the horizontal, tilt vertically upwards to vent the system Total weight: 37 kgPower supply
Operation is independent of the mains supply, thanks to high output rechargeable batteries with a capacity sufficient for up to two days' operation Input voltage charger: 100 - 240 V, 50 - 60 Hz Power consumption charger: < 100 VAMemory capacity
80 injection profiles can be defined and stored by the userFilling of injection syringes
Automatic filling via menu with volume input or
manual filling with variable speed Filling speed: 1 to 5 m l/s Optimised tube systems with check valves
Injection parameters
Maximum injection volume
- 200 ml
Partial injection volume
1 to 200 ml, programmable in 1 ml incrementsMaximum pressure
21 bar (305 psi), programmable from 5 to 21 bar in 1 bar increments (72.5 to 305 psi in 14.5 psi increments)Flow rate of the two injection units
0.1 - 10 ml/s, programmable in 0.1 ml/s increments Alternatively, input of flow rate or phase durationNumber of phases
1 to 6Injection delay
1 - 255 sPhase delay
1 - 255 sRemote control
Wireless touch-screen remote control Alternatively, display of injection parameters or pressure graphOptions
Use of prefilled syringes Alternatively, 65 ml/65 ml ELS
Product information and technical data
Product Image
Subject to technical alterations