Accutron MR - Mobile precision for magnet resonance imaging

Technical data

Application

  • Wireless, battery operated, fully digital user programmable double-head injector for contrast agent and physiological saline solution in the field of magnetic resonance imaging

Injector concept

  • Microprocessor controlled double-head injector with electric drive and two independent injection units
  • Multiphase, program controlled injection of contrast agent and physiological saline solution
  • Aluminium housing
  • Motor, control and rechargeable batteries are integrated into a single compact unit
  • Touch-screen control panel with different languages
  • Additional keypad for manual piston movement, starting and stopping the injection
  • Syringe holder with Easy Loading Syringe (ELS) 65 ml/200 ml
  • "Keep-Vein-Open" (KVO) feature
  • Automatic power saving mode
  • Non-stick syringe holder

Mechanical design

  • Swivelling injection units connected to a mobile support with a column stand
  • Operational angle of about 15° below the horizontal, tilt vertically upwards to vent the system
  • Total weight: 37 kg

Power supply

  • Operation is independent of the mains supply, thanks to high output rechargeable batteries with a capacity sufficient for up to two days' operation
  • Input voltage charger: 100 - 240 V, 50 - 60 Hz
  • Power consumption charger: < 100 VA

Memory capacity

  • 80 injection profiles can be defined and stored by the user

Filling of injection syringes

  • Automatic filling via menu with volume input or
    manual filling with variable speed
  • Filling speed: 1 to 5 m l/s
  • Optimised tube systems with check valves

Injection parameters

Maximum injection volume

  • 200 ml

Partial injection volume

  • 1 to 200 ml, programmable in 1 ml increments

Maximum pressure

  • 21 bar (305 psi), programmable from 5 to 21 bar in 1 bar increments (72.5 to 305 psi in 14.5 psi increments)

Flow rate of the two injection units

  • 0.1 - 10 ml/s, programmable in 0.1 ml/s increments
  • Alternatively, input of flow rate or phase duration

Number of phases

  • 1 to 6

Injection delay

  • 1 - 255 s

Phase delay

  • 1 - 255 s

Remote control

  • Wireless touch-screen remote control
  • Alternatively, display of injection parameters or pressure graph

Options

  • Use of prefilled syringes
  • Alternatively, 65 ml/65 ml ELS

Product information and technical data

Product Image


Subject to technical alterations