Application

  • Fully digital user programmable injector systems for contrast media in the field of angiography and/or computed tomography

Injector concept

  • Compact injector controlled by microprocessor
  • Activated by a brushless D.C. motor with electronic commutation
  • The head of the injector contains the mechanical drive mechanism and the control electronics
  • Integrated heated syringe holder with Easy Loading Syringe (ELS) 200 ml
  • Keypad with LC-Display on upper side of injector head

Mechanical Design

  • Injector head attached to mobile stand via pivotal arm
  • Operational angle of 15° below horizontal
  • Tilt vertically upwards to vent the system
  • Can be detached from the stand in a few movements
  • Handswitch to activate injection (5 m cable)
  • Total weight 27.5 kg

Power supply

  • Plug in power supply unit 230 V/50-60 Hz, optional 115 V/50-60 Hz
  • Max. power consumption: 375 VA
  • 5 m cable

Memory capacity

  • 100 programmes can be defined and saved by the user

Filling of injection syringe

  • Automatic filling via menue with volume input or
  • Manual filling with variable speed

Injection parameters

Maximum injection volume

  • 200 ml

Partial injection volume

  • 1 - 200 ml

Angiography conditions

Maximum pressure

  • 83 bar (1200 psi)

Flow rate

  • 0.1 - 30 ml/sec

Rise time

  • 0.1 - 10 sec

Volume

  • 1 - 200 ml

Number of phases

  • 1

Injection delay

  • 1 - 240 sec

X-ray delay

  • 1 - 240 sec

Number of protocols

  • 50
CT conditions

Maximum pressure

  • 21 bar (305 psi)

Flow rate

  • 0.1 - 10 ml/sec

Volume

  • 1 - 200 ml

Number of phases

  • 1 to 3

Injection delay

  • 1 - 240 sec

Phase delay

  • 1 - 240 sec

Scan delay

  • 1 - 240 sec

Number of protocols

  • 50

Accessories Injektron 82 HP

Options

  • Remote control
  • Wall or ceiling mounted support
  • Interface on request

Product information and technical data

Product Image

Subject to technical alteration